REF: 9516-163-50-ENG Rev: B1
ELI 250 / ELI 210 Service Manual
Manufactured by Mortara Instrument, Inc. Milwaukee U.S.A.
CAUTION: Use of this service manual is intended for qualified service personnel only.
CAUTION: Federal law restricts this device for sale to or on the order of a physician.
Page view 0
1 2 3 4 5 6 ... 110 111

Summary of Contents

Page 1 - REF: 9516-163-50-ENG Rev: B1

REF: 9516-163-50-ENG Rev: B1 ELI 250 / ELI 210 Service Manual Manufactured by Mortara Instrument, Inc. Milwaukee U.S.A. CAUTION

Page 3

ELI 250/210__________________________________________________________________________ A-12 Select F1 to sort the requests by patient name (ID, time

Page 4 - Other Important Information

__________________________________________________________________________Appendix A A-13 Custom ID Download Custom ID formats are uniquely defin

Page 5 - Warranty Information

ELI 250/210__________________________________________________________________________ A-14 “Waiting for Response”, “Connected”, and “Custom ID down

Page 6 - User Safety Information

__________________________________________________________________________Appendix B B-1 Appendix B ELI 250 Connectivity Options Modem Options Ther

Page 7

ELI 250/210__________________________________________________________________________ B-2 Xircom Modem Removal 1. Remove the top cover and the writ

Page 8

__________________________________________________________________________Appendix B B-3 Multitech Modem The Multitech modem is a separate board, ite

Page 9

ELI 250/210__________________________________________________________________________ B-4 3. Locate the 4 mounting screws , item # 45, for the modem

Page 10

__________________________________________________________________________Appendix B B-5 LAN Option The LAN, Local Area Network, option consists of t

Page 11 - 3 System Settings

ELI 250/210__________________________________________________________________________ B-6 Removal of the LAN Interface Board, item #58. 1. Open the

Page 12

__________________________________________________________________________Appendix B B-7 LAN Card Interconnect cables. The picture below shows the p

Page 13 - 5 Technical Description

ix Table of Contents 1 General Service Manual Purpose ...

Page 14

ELI 250/210__________________________________________________________________________ B-8 Wireless LAN Option The Wireless LAN option is a PCMCIA ca

Page 15 - 1 General

__________________________________________________________________________Appendix B B-9 Wireless LAN card (WLAN) Removal 1. Open the unit to access

Page 16

x Date Format ...3-9 Plot Format ...

Page 17 - 2 Maintaining and Cleaning

xi 5 Technical Description Introduction...

Page 18

xii Appendix B ELI 250 Connectivity Options Modem Options ...

Page 19

__________________________________________________________________________ Section 1 1-1 1 General Service Manual Purpose The purpose of this manual i

Page 20 - Interior Visual Inspection

ELI 250/210 _________________________________________________________________________ 1-2 Sterilizing this Product Do not sterilize this product or an

Page 21

____________________________________________________________________________Section 2 2-1 2 Maintaining and Cleaning CAUTION: Remove battery fuse

Page 22

ELI 250/210__________________________________________________________________________ 2-2 Cleaning and Inspecting Techniques This section contains in

Page 23

____________________________________________________________________________Section 2 2-3 Metallic and Plastic Parts Cleaning CAUTION Do not wipe o

Page 24 - 3-4

Copyright© 2004 by Mortara Instrument, Inc. 7865 N. 86th Street Milwaukee, Wisconsin 53224 This document contains confidential information that bel

Page 25

ELI 250/210__________________________________________________________________________ 2-4 Exterior Inspection Visually inspect the entire instrument

Page 26 - 3-6

____________________________________________________________________________Section 3 3-1 3 System Settings Chapter Purpose This chapter is intended

Page 27

ELI 250/210__________________________________________________________________________ 3-2 o Select Set Time/Date (number 3) from the appl

Page 28 - Configuration Page 1

____________________________________________________________________________Section 3 3-3 NOTES: Use Page (F4) to toggle through the configuration p

Page 29

ELI 250/210__________________________________________________________________________ 3-4 Summary of Configuration Menus Configuration Par

Page 30 - Configuration Page 2

____________________________________________________________________________Section 3 3-5 Append Unconfirmed Report, Reviewed by, Blank Screen Thre

Page 31

ELI 250/210__________________________________________________________________________ 3-6 Configuration Parameter Definition Configuration

Page 32 - Configuration Page 3

____________________________________________________________________________Section 3 3-7 3 + 3 Rhythm Lead 2 V1-V6, I, II, III, aVR, aVL, aV

Page 33

ELI 250/210__________________________________________________________________________ 3-8 Configuration Page 1 Software Version The di

Page 34 - Configuration Page 5

____________________________________________________________________________Section 3 3-9 (If necessary, you can use both the letter W and the letter

Page 35

i TECHNICAL SUPPORT AND SERVICE Headquarters Mortara Instrument, Inc. 7865 North 86th Street Milwaukee, WI 53224 Tel: 414.354.1600 Tel: 800.231.

Page 36

ELI 250/210__________________________________________________________________________ 3-10 Configuration Page 2 Flash Size Flash Size i

Page 37

____________________________________________________________________________Section 3 3-11 Filter The ECG plot frequency filter (or print filter) ca

Page 38

ELI 250/210__________________________________________________________________________ 3-12 The ELI 250 automatically analyzes ECGs and prin

Page 39

____________________________________________________________________________Section 3 3-13 This option defines the number of ECGs retrieved from an E

Page 40 - 3-20

ELI 250/210__________________________________________________________________________ 3-14 Configuration Page 4 Rhythm Formats Configurat

Page 41

____________________________________________________________________________Section 3 3-15 Configuration Page 6 (optional LAN connectivity) The ELI 2

Page 42

ELI 250/210__________________________________________________________________________ 3-16 Note: addresses are always entered as 4 sets of

Page 43

____________________________________________________________________________Section 3 3-17 Security (WEP) Wired Equivalent Privacy (WEP) is an encry

Page 44

ELI 250/210__________________________________________________________________________ 3-18 Access Configuration Menus for the ELI 210 Th

Page 45

____________________________________________________________________________Section 3 3-19 o Select Set Time/Date (number 3) from the application men

Page 46

ii Notices Manufacturer’s Responsibility Mortara Instrument, Inc., is responsible for the effects on safety and performance only if: • Assembly oper

Page 47 - Fuse Removal

ELI 250/210__________________________________________________________________________ 3-20 Summary of Configuration Menus Configuration Pa

Page 48

____________________________________________________________________________Section 3 3-21 Configuration Parameter Definition Configuration Screen

Page 49 - CAUTION

ELI 250/210__________________________________________________________________________ 3-22 Configuration Page 1 Software Version The d

Page 50

____________________________________________________________________________Section 3 3-23 This setting determines when the cardiograph will switch o

Page 51

ELI 250/210__________________________________________________________________________ 3-24 Configuration Page 2 Date Format This option

Page 52

____________________________________________________________________________Section 3 3-25 12 10 seconds of 12 leads in a 12 channel format 6+6

Page 53 - Keyboard/LCD Removal

ELI 250/210__________________________________________________________________________ 3-26 Configuration Page 3 Rhythm Formats Configurat

Page 54

___________________________________________________________________________ Section 4 4-1 4 Assembly/Disassembly of Unit CAUTION: Remove AC power

Page 55

ELI 250/210 _________________________________________________________________________ 4-2 Cover Assembly Removal / Installation From the standard op

Page 56

___________________________________________________________________________ Section 4 4-3 Flip unit over so that it is resting on the writer. Locate

Page 57

iii Warranty Information Your Mortara Warranty MORTARA INSTRUMENT, INC. (hereinafter referred to as “Mortara”) hereby warrants that Mortara products

Page 58

ELI 250/210 _________________________________________________________________________ 4-4 Open the writer door approximately 4 to 5 inches.

Page 59

___________________________________________________________________________ Section 4 4-5 Lift up and rotate the right side of the top cover to left a

Page 60

ELI 250/210 _________________________________________________________________________ 4-6 Slide top cover off the unit, rotating as needed to clear t

Page 61

___________________________________________________________________________ Section 4 4-7 Battery Removal Once the top cover is removed, the battery c

Page 62

ELI 250/210 _________________________________________________________________________ 4-8 Remove the LCD cable from the keyboard by 1) sliding the c

Page 63

___________________________________________________________________________ Section 4 4-9 The keyboard circuit board, item number 3 for the ELI 250 an

Page 64

ELI 250/210 _________________________________________________________________________ 4-10 When replacing the LCD, remove the old double sided tape a

Page 65

___________________________________________________________________________ Section 4 4-11 Writer Removal / Installation Located on the perimeter of

Page 66

ELI 250/210 _________________________________________________________________________ 4-12 Flip the unit over and remove the 4 screws, item 14, holdin

Page 67

___________________________________________________________________________ Section 4 4-13 Printhead Removal / Installation CAUTION The printhead

Page 68 - ELI 250/210 Block Diagram

iv User Safety Information Warning: Means there is the possibility of personal injury to you or others. Caution: Means there is the possibil

Page 69 - Front End Board

ELI 250/210 _________________________________________________________________________ 4-14 Carefully guide the printhead assembly out of the writer ch

Page 70

___________________________________________________________________________ Section 4 4-15 Writer Motor Removal / Installation Flip the writer assem

Page 71 - Specifications

ELI 250/210 _________________________________________________________________________ 4-16 To remove the writer motor remove the two screws, item num

Page 72 - Parts List

___________________________________________________________________________ Section 4 4-17 Printed Circuit Board Assembly Removal / Installation CA

Page 73

ELI 250/210 _________________________________________________________________________ 4-18 Once these cables are removed remove the 2 screws, item 45,

Page 74

___________________________________________________________________________ Section 4 4-19 CAUTION 1.) When installing a new board, do not replace

Page 75 - Troubleshooting

ELI 250/210 _________________________________________________________________________ 4-20 Locate the 4 screws holding the front-end to the chassis an

Page 76

____________________________________________________________________________Section 5 5-1 5 Technical Description Introduction The MORTARA ELI 250/

Page 77

ELI 250/210__________________________________________________________________________ 5-2 ELI 250/210 Block Diagram

Page 78 - Testing

____________________________________________________________________________Section 5 5-3 Main System Board The main system board or motherboard of

Page 79 - Writer Tests

v • Before attempting to use the device for clinical applications the operator must read and understand the contents of the manual and any documents

Page 80

ELI 250/210__________________________________________________________________________ 5-4 Patient Input Connector: Pin #: Name: Description: 1 V

Page 81 - Noise Test

____________________________________________________________________________Section 5 5-5 Specifications * Note – The ECG Storage capability of the

Page 82 - Transmitting Records

ELI 250/210__________________________________________________________________________ 5-6 Parts List Item Number Part Number Description 1 26025-0

Page 83

____________________________________________________________________________Section 5 5-7 Hardware Item Number Part Number Description 44 6020-430

Page 84 - Modem Test

ELI 250/210__________________________________________________________________________ 5-8

Page 85 - Application Menu

____________________________________________________________________________Section 6 6-1 6 Troubleshooting and Testing Introduction Since repair

Page 86 - Receive ECGs

ELI 250/210__________________________________________________________________________ 6-2 Unit turns on, no waveform is displayed but function key o

Page 87 - WLAN Test

____________________________________________________________________________Section 6 6-3 Optional Modem Transmit/Receive Problems If the unit does

Page 88 - Safety Tests

ELI 250/210__________________________________________________________________________ 6-4 Testing Power Supply Test 1.) Remove battery fuse and AC

Page 89 - Chapter Purpose

____________________________________________________________________________Section 6 6-5 Date / Time Test 1.) Turn the unit on. 2.) Press the F6, M

Page 90

vi Caution(s) • To prevent possible damage to the keypad, do not use sharp or hard objects to depress keys, only use fingertips. • Do not attemp

Page 91 - Modem Transmission

ELI 250/210__________________________________________________________________________ 6-6 Description of printout At the top of the writer test print

Page 92 - Country Code Country Code

____________________________________________________________________________Section 6 6-7 ECG Test 1.) Connect a simulator to the patient cable and c

Page 93 - WLAN Transmission

ELI 250/210__________________________________________________________________________ 6-8 Lead Fail Test Perform the lead fail test as follows: Con

Page 94 - LAN Transmission

____________________________________________________________________________Section 6 6-9 stored ECGs will remain in the directory until it becomes fu

Page 95

ELI 250/210__________________________________________________________________________ 6-10 The following screen appears during a batch transmission

Page 96

____________________________________________________________________________Section 6 6-11 Real Time ECG View Application Menu From the Applicati

Page 97

ELI 250/210__________________________________________________________________________ 6-12 TIP: “AT+GCI” will be highlighted – cursor is not presen

Page 98

____________________________________________________________________________Section 6 6-13 The LCD displays the following message: When the LCD dis

Page 99 - Patient Request List

ELI 250/210__________________________________________________________________________ 6-14 Channel Number: 6 WEP info (if used) Once this setup

Page 100

__________________________________________________________________________Appendix A A-1 Appendix A ELI 250 Connectivity Chapter Purpose This chapte

Page 101

vii • As defined by IEC 60601-1 and IEC 60601-2-25, the device is classified as follows: • Class I equipment or internally powered • Type CF appl

Page 102

ELI 250/210__________________________________________________________________________ A-2 An ECG may be listed in the directory but have a “deleted s

Page 103 - Modem Options

__________________________________________________________________________Appendix A A-3 After the transmission of your record(s), the Real-Time ECG

Page 104

ELI 250/210__________________________________________________________________________ A-4 From the Application Menu, simultaneously press ALT+SHIFT+

Page 105 - Multitech Modem

__________________________________________________________________________Appendix A A-5 Canada 20 Netherlands 7B Czech Republic 2E New Zealand 7E

Page 106

ELI 250/210__________________________________________________________________________ A-6 workstation. Also, it is required that your IT Manager set

Page 107 - LAN Option

__________________________________________________________________________Appendix A A-7 Caution: Possible damage to the cardiograph may occur i

Page 108

ELI 250/210__________________________________________________________________________ A-8 When the LCD displays the above message, the unit is read

Page 109

__________________________________________________________________________Appendix A A-9 ECGs are retrieved by ID number. Enter the desired ID and se

Page 110 - Wireless LAN Option

ELI 250/210__________________________________________________________________________ A-10 Under the heading “Site Selection”, the Request Code is di

Page 111

__________________________________________________________________________Appendix A A-11 When the Request List has been downloaded, you may select th