Modulus™ Microplate Multimode Reader Operating Manual Part Number 998-9300 Revision 1.6

Page 2

1 Instrument www.turnerbiosystems.com 5 NOTE: Do not connect anything except the injector system to the

Page 3 - Table of Contents

1 Instrument www.turnerbiosystems.com 6 1.2 Touch Screen Basics A built-in Windows-based PC with a custom interface enables user-frie

Page 4

1 Instrument www.turnerbiosystems.com 7 • Visible when a USB flash drive is detected • Icon turns dark when files have been saved to

Page 5

1 Instrument www.turnerbiosystems.com 8 • When viewing Absorbance results, the button toggles between the different data formats: OD,

Page 6

1 Instrument www.turnerbiosystems.com 9 maximum volume per well is 250 µl. Overfilling a well will result in spills and possible damage

Page 7

1 Instrument www.turnerbiosystems.com 10 1.3 HOME screen The HOME screen contains four options: • NEW PROTOCOL launches the wizard t

Page 8

1 Instrument www.turnerbiosystems.com 11 • NEW PROTOCOL allows the user to create new protocols. Touching this button launches the NEW

Page 9

1 Instrument www.turnerbiosystems.com 12 8. To save the modified protocol: 1. Touch the white text button with the protocol name. See

Page 10

1 Instrument www.turnerbiosystems.com 13 1.5.2 Well Selection To select or deselect a well, touch the well to toggle between green a

Page 11 - 1 Instrument

1 Instrument www.turnerbiosystems.com 14 Opening the instrument door while the Microplate Sample Tray is in motion will cause it to st

Page 12

Turner BioSystems, Inc. and its suppliers own the written and other materials in the Modulus Microplate Reader operating manual

Page 13

1 Instrument www.turnerbiosystems.com 15 Use the DELETE button to permanently remove unwanted results from the re

Page 14

1 Instrument www.turnerbiosystems.com 16 1.7 Transferring Data Results files are transferred to a PC or Mac via a USB flash drive usi

Page 15 - 1.4 Protocol Management

1 Instrument www.turnerbiosystems.com 17 1.8 Repeat Runs Each detection modality has the option to do repeat runs. Injectors cannot b

Page 16

2 Luminescence www.turnerbiosystems.com 18 2 Luminescence 2.1 General Information • The Luminometer detection wavelength range is 350 to 65

Page 17 - 1.5 Instrument Control

2 Luminescence www.turnerbiosystems.com 19 2.4 Starting a Run Easy-to-use protocol wizards can be accessed from the HOME screen to set up the

Page 18

2 Luminescence www.turnerbiosystems.com 20 2.5 Kinetic In kinetic mode, the instrument performs multiple reads of a single well with

Page 19

2 Luminescence www.turnerbiosystems.com 21 2.7 Using a Standard Light Plate The optional Standard Light Plate provides a quick way to

Page 20

2 Luminescence www.turnerbiosystems.com 22 2.8 Luminescence Data Luminescence data is displayed in Relative Luminescence Units (RLU).

Page 21 - 1.7 Transferring Data

www.turnerbiosystems.com 23 3 Fluorescence 3.1 General Information • Microplate Recommendations: - Black plates are highly recommended. -

Page 22 - 1.8 Repeat Runs

3 Fluorescence www.turnerbiosystems.com 24 3.3 Starting a Run Easy-to-use protocol wizards can be accessed from the HOME screen to set

Page 23 - 2 Luminescence

Table of Contents 1 Instrument ...

Page 24 - 2.4 Starting a Run

3 Fluorescence www.turnerbiosystems.com 25 3.4 Switching Fluorescence Optical Kits 1. Close the instrument door by touching the DOOR

Page 25 - 2.6 Kinetic Parameters

3 Fluorescence www.turnerbiosystems.com 26 5. Hand tighten the 2 captive screws using the Allen Wrench (size 7/16) provided with the ki

Page 26

www.turnerbiosystems.com 27 4 Absorbance 4.1 General Information • Four installed filters and two customizable filter holders are included on

Page 27 - 2.9 Luminescence Module

4 Absorbance www.turnerbiosystems.com 28 4.2.1 Customizable Absorbance Paddle Two absorbance paddles are included for customization. T

Page 28 - 3 Fluorescence

4 Absorbance www.turnerbiosystems.com 29 3. Touch the white text button to select a protocol or modify the displayed parameters. See Fi

Page 29 - 3.3 Starting a Run

4 Absorbance www.turnerbiosystems.com 30 Wells marked with a Ref were designated as Reference wells. Reference value of each wavelength

Page 30 - 3.6 Fluorescence Module

4 Absorbance www.turnerbiosystems.com 31 9. To ensure the instrument detects the newly installed module, verify the ABSOR icon appears

Page 31

www.turnerbiosystems.com 32 5 Injector System 5.1 Injector System Components Unpack the Modulus™ Microplate Reader injector system carefully. V

Page 32 - 4 Absorbance

5 Injector System www.turnerbiosystems.com 33 1. Power off the Modulus™ Microplate Reader. 2. Place the Inject

Page 33 - 4.3 Starting a Run

5 Injector System www.turnerbiosystems.com 34 6. Manually open the instrument door and hold it down. Insert the injector tip into the

Page 34 - 4.5 Absorbance Data

Table of Contents 2.9 Luminescence Module ...

Page 35 - 4.6 Absorbance Module

5 Injector System www.turnerbiosystems.com 35 11. Pinch the lever on the back of the inlet tube holder to clip the stainless steel tube

Page 36

5 Injector System www.turnerbiosystems.com 36 NOTE: The reagent container holder can accommodate four size containers: A round flat bott

Page 37 - 5 Injector System

5 Injector System www.turnerbiosystems.com 37 5.3.1 Priming Injectors 1. Place the reagent container into the reagent holder. 2.

Page 38

5 Injector System www.turnerbiosystems.com 38 2. Place the container of 70% ethanol on the bottle holder. The stainless steel inlet tub

Page 39

5 Injector System www.turnerbiosystems.com 39 5.8 Removing or Replacing Inlet and Outlet Plastic Tubing Replacement tubing assembly is availab

Page 40

www.turnerbiosystems.com 40 6 Accessories 6.1 Curve Fitting Data Analysis Software (9300-970) Curve Fitting Data Analysis software is easy-to-us

Page 41

6 Accessories www.turnerbiosystems.com 41 • When an external PC has control of the instrument, the Modulus™ Microplate Reader touch s

Page 42 - 5.4 Cleaning the Injectors

6 Accessories www.turnerbiosystems.com 42 3. Power on the Modulus™ Microplate Reader. 4. The FOUND A NEW HARDWARE wizard will appe

Page 43 - 5.6 Replacing Injector Tips

6 Accessories www.turnerbiosystems.com 43 6.2.7 Connecting to a PC Using an RS-232 Serial Cable 1. Power on the instrument and PC.

Page 44

6 Accessories www.turnerbiosystems.com 44 6.3.3 Changing the Battery 1. Remove the two Phillips head screws. 2. Remove the batter

Page 45 - 6 Accessories

Table of Contents 6 Accessories ...

Page 46 - 6 Accessories

www.turnerbiosystems.com 45 7 Tools From the INSTRUMENT CONTROL screen, touch the TOOLS button to manage various aspects of the instrument. 7.1

Page 47

7 Tools www.turnerbiosystems.com 46 7.2 Event Log The event log can be downloaded onto a USB flash drive in a .TXT format. Insert

Page 48

7 Tools www.turnerbiosystems.com 47 7.6 Updating Firmware Perform a firmware update only when instructed by Turner BioSystems Tech

Page 49

www.turnerbiosystems.com 48 8 Instrument Maintenance 8.1 General Instrument Care The instrument must be on a level surface to operate properly.

Page 50 - 7 Tools

8 Instrument Maintenance www.turnerbiosystems.com 49 8.3 Touch Screen Care Do not use any sharp object such as a pen, pencil, sty

Page 51 - 7.5 Updating Software

www.turnerbiosystems.com 50 9 Troubleshooting 9.1 Table of Error Messages Error Message Description Resolution Start Problem Instrument busy T

Page 52 - 7.6 Updating Firmware

9 Troubleshooting www.turnerbiosystems.com 51 Injector 2 not installed The protocol requires addition of reagent using Injector 2 but it is not d

Page 53 - 8 Instrument Maintenance

9 Troubleshooting www.turnerbiosystems.com 52 9.2 Table of Common Problems Symptom Possible Cause Resolution Injections sputter, dr

Page 54 - 8.3 Touch Screen Care

9 Troubleshooting www.turnerbiosystems.com 53 9.3 Commonly Asked Questions 9.3.1 PC Connectivity Q: When I connect the Modulus™ M

Page 55 - 9 Troubleshooting

www.turnerbiosystems.com 54 APPENDIX Appendix A - Specifications General Specifications • Detection Modes: Luminescence, Fluorescence, and Abs

Page 56

Table of Contents APPENDIX ...

Page 57

APPENDIX www.turnerbiosystems.com 55 Injector System Specifications • Number of Injectors: single or dual injectors • Dispense Volume

Page 58

APPENDIX www.turnerbiosystems.com 56 Appendix B - Warranty and Obtaining Service General Warranty Information Turner BioSystems warra

Page 59 - APPENDIX

APPENDIX www.turnerbiosystems.com 57 Prior correspondence is necessary to: • Eliminate minor problems, easily handled on-site in your

Page 60

APPENDIX www.turnerbiosystems.com 58 Appendix C - Ordering Information To get a price quote or place an order call the Turner BioSyst

Page 61

APPENDIX www.turnerbiosystems.com 59 Part Number Description Includes Replacement Part 159-9100 Power supply Brick 24V, 60W, 2.5A 046-

Page 62

APPENDIX www.turnerbiosystems.com 60 Appendix D - How to Contact Us Turner BioSystems, Inc. 645 North Mary Avenue Sunnyvale, CA 94085

Page 63

www.turnerbiosystems.com 61

Page 65

645 N. Mary Avenue • Sunnyvale, CA 94085 Phone: (408) 636-2400 •Toll Free: (888) 636.2401 • Fax: (408) 737-7919 www.turnerbiosystems.com

Page 66

www.turnerbiosystems.com 2 1 Instrument 1.1 Getting Started 1.1.1 Description Thank you for purchasing the Turner BioSystems, Inc. Modulus™ Mic

Page 67

www.turnerbiosystems.com 3 1.1.3 Unpacking and Inspection Upon receiving the Modulus™ Microplate Reader, inspect it carefully for any damage to the

Page 68 - Sunnyvale, CA 94085

1 Instrument www.turnerbiosystems.com 4 3. Plug the power line cord into the power connector on the back panel of the instrument. Plug