SParameter setting support software for the IV seriesIV-S30SP Version 3.0Produced in Sep. 2002Instruction ManualModel nameR<Applied models>- I
Pref-3Pref.Preface[4] Product componentsOne CD-ROMOne Instruction manualOne RS-232C communication cable (1.5 m)[Connector --- 9-pin D-sub male (IV-S30
Chapter 4: Document Creation4-24IV-S20From the previous page(4) Starting up the document creation screenClick the "Creates a document" item
4-34Chapter 4: Document CreationIV-S204-2 Document creation (file data)This section describes the procedures for creating a document from the paramete
Chapter 4: Document Creation4-44IV-S20[Example of document creation]Shown below is an example which a document file (with ".csv" extension)
5-15Chapter5: Data CollectionIV-S20Chapter 5: Data CollectionYou can transmit the measurement result data from the IV-S20 to a personal computer via a
5-25Chapter 5: Data CollectionIV-S20From the previous page(4) File selectionSelect the "Open" item from the "File" menu on the &qu
5-35Chapter5: Data CollectionIV-S20From the previous page(6) Execute a data collectionClick on the "START" button.- The data collection from
5-45Chapter 5: Data CollectionIV-S20[Data collection example]Shown below is an example in which the graph creation feature in Excel is used to make a
6-16Chapter 6: Reading/Writing Parameters and ImagesIV-S20Chapter 6: Reading/Writing Parameters and ImagesSetting parameters and images (display image
6-26Chapter 6: Reading/Writing Parameters and ImagesIV-S20Below describes the operation procedures to save (IV -> File) the IV-S20 setting paramete
6-36Chapter 6: Reading/Writing Parameters and ImagesIV-S206-2 Verifying setting parametersSelect the "Verify" item from the "IV data se
Setup-1SetSetupFace the CD-ROM label side up.SetupThe installation process that places a program on a computer and makes it ready for use is referred
6-46Chapter 6: Reading/Writing Parameters and ImagesIV-S206-3 Loading setting parametersSelect the "Load" item from the "IV data settin
6-56Chapter 6: Reading/Writing Parameters and ImagesIV-S206-4 Reading/writing images[1] Reading images (display images and messages)Select the "R
6-66Chapter 6: Reading/Writing Parameters and ImagesIV-S20From the previous page2. Click the "OK" button on the "Read display image (CA
6-76Chapter 6: Reading/Writing Parameters and ImagesIV-S20[2] Writing images (load images)On the "RUN MENU SETTINGS" menu of the IV-S20, sel
6-86Chapter 6: Reading/Writing Parameters and ImagesIV-S20ClickFrom the previous page(3) Loading display images1. Click the "Write image" →
Chapter 7: Command Test7-17IV-S20Click hereOn the "MAIN OPS MENU", move the cursor to SET-SCRN, and press the SET key.- On the [SYSTEM SETUP
Chapter 7: Command Test7-27IV-S20From the previous page(4) Setting the conditions for the command testIn the "Command test" dialog box, set
Chapter 7: Command Test7-37IV-S20From the previous page(6) Saving/loading files- Saving the data received to a file1. Click on the "Select a file
8-110Chapter 8: Upgrade Version8IV-S20Continued on the following pageSet key (SET)Left key (\)[IVS20 VERSION UP MENU]1SYSTEM RECEIVE RUN2SYSTEM TR
8-210Chapter 8: Upgrade Version8IV-S20From the previous page(4) File selectionClick on the "Select a file" item on the "Version up"
Setup-2SetSetup5. Click the "Next" button. The following dialog box will appear.6. When you do not need to change the folder (directory) whe
9-111Chapter 9: Additional Descriptions9IV-S20Chapter 9: Additional DescriptionsThis chapter describes additional description; copy to the clipboard,
9-211Chapter 9: Additional Descriptions9IV-S20[3] Changing the image brightnessYou can set the displayed image brightness to "Normal" or &qu
E-1EError codeIf an error occurs and an error code appears while operating the IV-S30SP, use the error code list below fordetails about the error.For
Setup-3SetSetup7. By clicking on the "Next" button, the "Select Program Folder" dialog box will appear.8. When you do not want to
Setup-4SetSetup10. To start this program, bring up the start menu in Windows98, and then select "Programs" - "Sms" -and "IVS3
Setup-5SetSetup· You can also stop the program by clicking on the " " button in the upper right corner of the window.- The confirmation dial
Setup-6SetSetup[3] Installation of a USB driver (version for the IV-S30/C35M)1. Turn ON the power to the controller (IV-S33MX etc., except IV-S30J).2.
Setup-7SetSetup3. Click on the [Next] button. The display will change to the screen for selecting a search method. Select"Search for the best dri
Setup-8SetSetup6. To make sure that the setting is completed correctly, select [System] from [Control Panel] and exam-ine the device tree in the Devic
Version for the IV-S30/C35M
Thank you for purchasing the IV-S30SP, which is parameter setting support software for the IV series.Read this manual thoroughly to understand the fea
1-1Chapter 1: Menu Organization1IV-S30/C35MChapter 1: Menu OrganizationAfter starting this software (for the IV-S30/C35M), the screen shown below will
1-2Chapter 1: Menu Organization1IV-S30/C35M(1) File(2) EditCut Ctrl + X Copy the image displayed on the IV-S30SP to the clipboard and delete it
1-3Chapter 1: Menu Organization1IV-S30/C35M(3) View (display)(4) IVS30 communicationToolbars Standard Select whether or not to display the standard
1-4Chapter 1: Menu Organization1IV-S30/C35M(5) Tools(6) Window(7) HelpContents Display the table of contents for the Help files.Contents and Index D
2-1Chapter 2: Optional Settings2IV-S30/C35MChapter 2: Optional SettingsWhen you communicate between the personal computer and the IV-S30/C35M, you mus
2-2Chapter 2: Optional Settings2IV-S30/C35MË USB communicationSelect "USB" for the "COMM method" and specify the other communicati
2-3Chapter 2: Optional Settings2IV-S30/C35M2-2 Transfer conditionsSelect "IV-S30/C35M transfer conditions" from the "Option" dialo
2-4Chapter 2: Optional Settings2IV-S30/C35M2-3 Upgrade conditionsSelect "Software upgrade conditions" in the "Option" dialog box a
3-1Chapter 3: Parameter Settings3IV-S30/C35MChapter 3: Parameter settingsSet object type conditions (Measurement conditions, Evaluation conditions, Nu
3-2Chapter 3: Parameter Settings3IV-S30/C35M Setting the measurement conditions: When measurement 0 and camera 1 (position deviation measure-ment) is
ItemModelCPUOperating systemMemoryHard diskDisplayCD-ROM driveRS-232C portMousePrinterSpecificationIBM PC/AT or compatible machinesPentium 90 MHz mini
3-3Chapter 3: Parameter Settings3IV-S30/C35M- The "Measure position deviation" setting screen will appear.
3-4Chapter 3: Parameter Settings3IV-S30/C35M3-1 Setting the measurement conditionsSelect the "Measure position deviation" tab and double-cli
3-5Chapter 3: Parameter Settings3IV-S30/C35M3-2 Setting the evaluation conditionsSelect the "Object type condition" tab and double-click on
3-6Chapter 3: Parameter Settings3IV-S30/C35M3-3 Setting the numerical calculation conditionsSelect the "Object type conditions" tab and doub
3-7Chapter 3: Parameter Settings3IV-S30/C35MClick on each of the individual setting sections 1 to 5 below, on the numerical calculation setting table.
3-8Chapter 3: Parameter Settings3IV-S30/C35MDescribed below is the process used to set the calculation result N0 for the setting example shown on thep
3-9Chapter 3: Parameter Settings3IV-S30/C35MFrom the previous page(3) Setting the upper & lower limitsSpecify the upper limit for evaluating the o
3-10Chapter 3: Parameter Settings3IV-S30/C35M3-4 Setting the output conditionsSelect the "object type conditions" tab and double-click on th
3-11Chapter 3: Parameter Settings3IV-S30/C35M[Example of settings]Described below is the process used to create the example above.1. Enter "0&quo
3-12Chapter 3: Parameter Settings3IV-S30/C35M7. Specify the output number for the output signal, and the click on the "Input" (enter) button
第1章 メニュー構成第2章 通信設定第3章 品種別条件の設定第4章 ドキュメント作成第5章 データ収集第6章 パラメータ、画像の読出/書込第7章 コマンドテスト第8章 バージョンアップ第9章 その他の機能PrefaceSetup Version for the IV-S30/C35MChap
3-13Chapter 3: Parameter Settings3IV-S30/C35M3-5 Setting the distance and angle conditionsSelect the "Object type conditions" tag and double
3-14Chapter 3: Parameter Settings3IV-S30/C35M2. Specify the auxiliary conditions in the "Distance and angle conditions" dialog box.3. Specif
4-14IV-S30/C35MChapter 4: Reading/Writing Parameters and ImagesChapter 4: Reading/Writing Parameters and ImagesYou can save the parameter settings and
Chapter 4: Reading and Writing Parameters and Images4-24IV-S30/C35M4-2 WriteWrite the parameter settings and images stored into a personal computer to
4-3Chapter 4: Reading and Writing Parameters and Images4IV-S30/C35M4-3 VerifyVerify that the parameter settings and images are the same in the persona
Chapter 4: Reading and Writing Parameters and Images4-44IV-S30/C35M4-4 InitializationInitialize all parameters in the IV-S30/C35M.1. Select the "
4-5Chapter 4: Reading and Writing Parameters and Images4IV-S30/C35M4-5 Self diagnosisStart the self-diagnosis function in the IV-S30/C35M.1. Select th
5-15Chapter 3: File OperationsIV-S30/C35MChapter 5: File operationsSave the IV-S30/C35M's parameter settings and images in a file or load the sam
5-25Chapter 3: File OperaionsIV-S30/C35MË Save As projectSave all of the files currently open, which have any of the extensions described in the &quo
6-1Chapter 6: Document Creation6IV-S30/C35MClick hereContinued on the following pageChapter 6: Document CreationYou can automatically create a spreads
Table of contentsPreface ... Pref-1 to 3[1] Fe
Chapter 6: Document Creation6-26IV-S30/C35MFrom the previous page(4) Create a document from a project file.Click on the "Open project file"
6-3Chapter 6: Document Creation6IV-S30/C35M6-2 Document creation (IV-S30/C35M parameter loading)This section describes the procedures for creating a d
Chapter 6: Document Creation6-46IV-S30/C35M[Example of document creation]Shown below is an example file document (with ".csv" extension) tha
7-1Chapter 7: Data Collection7IV-S30/C35MContinued on the following page[OBJECT TYPE I/O]1TRIG CCD START NO YES(BIN AVG-GRAYS GRAY-SRC)2CCD TRIG CO
7-2Chapter 7: Data Collection7IV-S30/C35MFrom the previous page3. Set the "SERIAL OUTPUT" or "OUT I/F (PARAL.)" item on the [I/O C
7-3Chapter 7: Data Collection7IV-S30/C35MClick hereClick hereContinued on the following pageClick hereFrom the previous page(2) Starting the data coll
7-4Chapter 7: Data Collection7IV-S30/C35MContinued on the following pageClick hereFrom the previous page[Sampling data]- Measurement result data: Coll
7-5Chapter 7: Data Collection7IV-S30/C35MClick here121. Click on the "Cancel" button.- The data collection will stop.2. Click on the "C
7-6Chapter 7: Data Collection7IV-S30/C35M2500025200254002560025800260002620026400266002680027000272002740027600278002800028200284002860028800290002920
8-1Chapter 8: User Menu Editor108IV-S30/C35MChapter 8: User Menu EditorThis function lets you create and modify the MAIN OPS MENU screen and all of th
Chapter 8: User Menu Editor (IV-S31M/S32M/S33M) ... 8-1to 8-98-1 How to start the editor 8-18-2 Read 8-28-3 Write 8-
8-2Chapter 8: User Menu Editor108IV-S30/C35M8-2 ReadRead in the menu you want to edit, or select the MAIN OPS MENU screen, from the IV-S33M etc.1. Cli
8-3Chapter 8: User Menu Editor108IV-S30/C35M8-4 Initialize the user menus1. Click on the "Data Transfer" item in the "File" menu (
8-4Chapter 8: User Menu Editor108IV-S30/C35M8-5 Editing TitlesThe word "title" refers to the "MEAS SHAPE (MDL0)" text shown inside
8-5Chapter 8: User Menu Editor108IV-S30/C35M- The "Find Title list" dialog box will appear.2. Enter the phrase you want to find. Then click
8-6Chapter 8: User Menu Editor108IV-S30/C35M8-7 Menu editing"Character strings" refers to any part of the complete line of text in the examp
8-7Chapter 8: User Menu Editor108IV-S30/C35M8-8 Create a user menu1. Click on the "Create user menu" item in the "File" menu.- The
8-8Chapter 8: User Menu Editor108IV-S30/C35M8-9 Create a new MAIN OPS MENU screen1. Click on the "Create MAIN OPS MENU menu" item from the &
8-9Chapter 8: User Menu Editor108IV-S30/C35M3. Select the area that you want to assign to the MAIN OPS MENU screen from the area list. To add theitem,
Chapter 9: Command Test9-1119IV-S30/C35MContinued on the following pageClick hereChapter 9: Command TestThe command test function is used to communica
Chapter 9: Command Test9-2119IV-S30/C35MFrom the previous page(4) Setting the conditions for the command testIn the "Command test" dialog b
6-4 Reading/writing images 6-5[1] Reading images (display images and messages) 6-5[2] Writing images (load images) 6-7Chapter 7: Command Test ...
Chapter 9: Command Test9-3119IV-S30/C35MFrom the previous page(6) Saving files- Saving the data received to a file1. Click on the "Open" it
10-1Chapter 10: Upgrade Version10IV-S30/C35MContinued on the following pageSet key (SET)Left key (\)"XX" will vary depending on the models,
10-2Chapter 10: Upgrade Version10IV-S30/C35MFrom the previous page(4) File selectionClick on the "Version upgrade" item on the "IVS30 c
11-1Chapter 11: Additional Descriptionsはじめにセットアップ11IV-S30/C35MChapter 11: Additional DescriptionsThis chapter describes other functions such as Printi
11-2Chapter 11: Additional Descriptions11IV-S30/C35M[3] ZoomYou can change the display magnification of the image by changing the zoom setting.(Zoom s
11-3Chapter 11: Additional Descriptionsはじめにセットアップ11IV-S30/C35M[5] SMS Web pageIf your personal computer is ready to connect to the Internet, the defau
11-4Chapter 11: Additional Descriptions11IV-S30/C35M[6] Memory card (IV-C35M)When the IV-C35M is used, you can read, save, or delete data from a flash
Version for the IV-S20
1-1Chapter 1: Menu Organization1IV-S20Chapter 1: Menu OrganizationAfter starting this software (for the IV-S20), the screen shown below will appear. T
1-2Chapter 1: Menu Organization1IV-S20(1) File(2) EditOpen an existing image file. (*.bmp)(Description)Create a new file.New Ctrl + NOpen image Ctrl
Pref-1Pref.PrefacePrefaceThis software (IV-S30SP) is a parameter setting support software used to specify the parameters for the com-pact image sensor
1-3Chapter 1: Menu Organization1IV-S20File - Set screenSelect an IV-S20 setting data file to treat.(Description)LoadSaveLoad the setting parameters fr
1-4Chapter 1: Menu Organization1IV-S20(4) Object type COND(Description)Object type NO. (0 - 15)Set and change object type number.Title registrationSet
1-5Chapter 1: Menu Organization1IV-S20(5) Version up(6) Window(7) Help(Description)Select a fileTransfer data Without INITTransfer data and INIT All d
2-12Chapter 2: Set CommunicationIV-S20Chapter 2: Set CommunicationWhen you communicate between the PC and the IV-S20, you must specify the "Set c
3-13Chapter 3: Setting Object Type ConditionsIV-S20Chapter 3: Setting Object Type ConditionsSet object type conditions (Measurement conditions, Evalua
3-23Chapter 3: Setting Object Type ConditionsIV-S20Edge detectionconditions set-ting items.3-1 Setting the measurement conditionsClick each measuring
3-33Chapter 3: Setting Object Type ConditionsIV-S203-2 Setting the evaluation conditionsClick the "EVALUATION CONDITION" from each measuring
3-43Chapter 3: Setting Object Type ConditionsIV-S20Click3-3 Setting the numerical calculation conditionsClick the "NUMERIC CALC. COND" from
3-53Chapter 3: Setting Object Type ConditionsIV-S20Click on each of the individual setting sections 1 to 5 below, on the numerical calculation setting
3-63Chapter 3: Setting Object Type ConditionsIV-S20Described below is the process used to set the calculation result N01 for the setting example shown
Pref-2Pref.Preface[2] Operating environmentThis software runs on any computer, which is equipped with the following operating environment.· Windows is
3-73Chapter 3: Setting Object Type ConditionsIV-S20Click hereFrom the previous pageDouble-click hereIndicated1 Click the " item to select thenumb
3-83Chapter 3: Setting Object Type ConditionsIV-S203-4 Setting the output conditionClick the "OUTPUT CONDITIONS" from each measuring program
3-93Chapter 3: Setting Object Type ConditionsIV-S20[Example of settings]Described below is the process used to create the example above.1. Click the c
3-103Chapter 3: Setting Object Type ConditionsIV-S206. Click on the [AUXILIARY C] in the "OUTPUT SIGNAL" area.- The output signal and the au
3-113Chapter 3: Setting Object Type ConditionsIV-S20Click3-5 Operation example (positional deviation measurement)This section describes the operation
3-123Chapter 3: Setting Object Type ConditionsIV-S20From the previous page- Setting screen of positional deviation measurement(6) Reading display imag
3-133Chapter 3: Setting Object Type ConditionsIV-S20From the previous page- An example of reading display image(7) Mode settings1. Specify a register
3-143Chapter 3: Setting Object Type ConditionsIV-S20The reference imageframe for the model 0The reference imageframe for the model 1The search areafra
3-153Chapter 3: Setting Object Type ConditionsIV-S20From the previous page2. Click the "REG." button. Then click the "DISP" button
4-14Chapter 4: Document CreationIV-S20Chapter 4: Document CreationYou can automatically create a spreadsheet (table of the current parameters) using E